Safety starts before your first session.

Intended use, contraindications and supervision according to IFU RS-200-1A v3.1. This is device information; individual assessment for treatment is conducted by a specialist.

Intended use and age range

According to IFU, the device is intended to slow the progression of myopia in children aged 3-16. It is intended for use at home, according to a specific session schedule.

The diagnosis of myopia must be confirmed before use. A child's age alone does not mean that he or she can use the device.

Source: IFU RS-200-1A, v3.1, 18/07/2023 , pp. 9, 16. The description does not replace the instructions supplied with the device.

Assessment, parent and specialist

Before first use, a specialist should assess whether the purpose and conditions of use of the device correspond to the child's situation. The parent or guardian must know the instructions and supervise the sessions.

IFU specifically highlights the need to instruct children aged 3–8 years. Users with color vision disorders also require supervision from a parent or guardian.

Do not start therapy or change medications based on the site's content.

Source: IFU RS-200-1A, v3.1, 18/07/2023 , pp. 9–10, 16. The description does not replace the instructions supplied with the device.

Contraindications from the instructions

IFU v3.1 lists the following groups of contraindications:

  • strabismus;
  • binocular vision abnormalities;
  • ophthalmic or other systemic abnormalities;
  • mydriasis, including after atropine, cyclopentolate or tropicamide;
  • retinal diseases in children - the manual lists, among others: retinopathy of prematurity, retinal detachment, juvenile macular degeneration, retinoblastoma and uveitis.

RLRL is not used concomitantly with atropine.

The instructions exclude use when pupils are dilated, including after atropine. Do not stop atropine or other medicines on your own. The treating specialist determines suitability for therapy and the subsequent care plan.

Source: IFU RS-200-1A, v3.1, 18/07/2023 , p. 9. The description does not replace the instructions supplied with the device.

Warnings and conditions of use

  • The source is a class 3R laser. Use the device only as specified in the instructions.
  • Please remove your glasses and contact lenses before your session.
  • Don't extend the session. Keep a minimum break of 4 hours and a maximum of 10 sessions per 7 days.
  • Do not disassemble the device and do not use unauthorized parts.
  • Disconnect the power before cleaning. Do not allow the interior to flood.

The provided IFU distinguishes between the expected life of the device (5 years) and the maximum period of continuous safe use (2 years). Don't treat the period of use as consent to five years of therapy. Continued use requires reference to the current device-specific instructions and expert judgment.

Source: IFU RS-200-1A, v3.1, 18/07/2023 , pp. 9, 15–17. The description does not replace the instructions supplied with the device.

Symptoms and stopping use

The instructions describe possible glare, temporary flash blindness and afterimages. They also mention light sensitivity, irritation and a burning sensation.

Stop using the device and contact a specialist

If afterimages persist, or pain, visual disturbance, irritation or another concerning symptom occurs, do not continue with further sessions until the cause has been assessed.

In the public FAQ, the manufacturer indicates an afterimage lasting more than 5 minutes as a reason for consultation. This is not an encouragement to wait if symptoms are severe.

Source: IFU RS-200-1A, v3.1, 18/07/2023 , pp. 16–17. The description does not replace the instructions supplied with the device.

Addendum: public FAQ Eyerising International .

Monitoring goes beyond recording sessions

The session recording system does not measure the eye's refraction or axial length. Ophthalmological check-ups and assessment of progress are carried out by a specialist.

Work with your provider to establish a testing schedule, a method of reporting symptoms, and criteria for continued use. Do not replace the visit with an evaluation of the device screen itself.

Reporting a problem or incident

Please consult a specialist regarding your symptoms. Also report a technical problem or event related to the device to Euromedic Poland, providing the model and serial number as well as the time and description of the event.

Do not send medical records via a regular form or first message. Agree with the staff on a secure method of transferring data, if needed.

Contact Euromedic Poland

Documents and literature

Sources checked on 10/09/2026. The version of local PDF is provided explicitly; We do not mark it as the latest manual for all markets.

Let's talk about therapy

Specific questions.
Direct contact.

Therapy conditions, documentation and operation of the device.

+48 512 121 252

Monday–Friday, 9:00–17:00

Opens in a new tab.